Overall Performance
|
|
|
Equipment type
|
Fully automatic discrete, STAT priority
|
|
Analysis rate
|
Constant speed 200T/H (without ISE)
|
|
Test principle
|
Colorimetry, turbidimetry
|
|
Analysis method
|
End-point, 2 kinetics, fixed-time, etc. support single/double wavelength, linear and non-linear calibration
|
|
Sample System
|
|
|
Sample Tray
|
42 sample positions, random access of all samples
|
|
Sample Volume
|
2-50ul, 0.1ul increment
|
|
Sample probe
|
Liquid level detection, the remaining detection, collision protection etc.
|
|
Sample Containers
|
Sample cups, Micro tube, Original blood-collection tube, etc.
|
|
Reagent System
|
|
|
Reagent Tray
|
25 reagent positions
|
|
Reagent Volume
|
R1:25-300ul, R2:10-150ul, 0.1ul increment
|
|
Reagent Probe
|
Liquid level detection, the remaining detection, collision protection etc.
|
|
Reagent Refrigeration
|
4-15℃, independent power supply, 24-hour non-stop refrigeration
|
|
Reaction System
|
|
|
Reaction Tray
|
Rotating tray, 120 cuvettes
|
|
Cuvette
|
Reusable high quality cuvette, Optical diameter 6mm
|
|
Reaction Volume
|
180-500ul
|
|
Incubation system
|
Bi-directional liquid-air device
|
|
Temperature
|
37℃±0.1℃ with real-time display
|
|
Optical System
|
|
|
Light source
|
12V35W halogen lamps, over 2000 hours
|
|
Photoelectron road
|
after spectrophotometry
|
|
Wavelength
|
340,405,450,510,546,578,630,700nm
|
|
Absorbance range
|
0-3.500Abs
|
|
Resolution
|
0.0001Abs
|
|
Operating System
|
|
|
PC operating system
|
Windows XP or above
|
|
Analysis control software
|
English version graphical operating software
|
|
Report printing
|
Report formats support the user-defined mode, QC and state information etc.
|
|
System connection
|
RS-232
|
|
Working Conditions
|
|
|
Power supply
|
AC 220V±22V, 50Hz±1Hz, power 1.5KVA
|
|
Ambient Temperature
|
15-30 ℃
|
|
Humidity
|
35-80%
|
|
Water consumption
|
5L/Hour
|
|
Dimension
|
Bench top: 750mm*520mm*600mm
|
|
Weight
|
45KG
|